  1. The duration of transient absorption spectroscopy measurements typically limits the types of systems for which the excited state dynamics can be measured. We present a single-shot transient absorption (SSTA) instrument with a spatially encoded 60 ps time delay range and a 100 nm spectral range that is capable of acquiring a transient spectrum in 20 s. We describe methods to spatially overlap the flat-top pump and probe beams at the sample plane, calibrate the spatially encoded time delay, and correct for non-uniform excitation density. SSTA measurements of organic materials in solution and film demonstrate this technique. 